Steam sterilizer GK-100-3. For sterilization with saturated water steam under pressure of medical products.
Steam semi-automatic sterilizer GK-100-3 with vacuum drying, economical, easy and convenient to use, safe for the environment.
It is intended for sterilization with water-saturated steam under pressure of medical products made of metals, glass, rubber, ligature suture material, textile products, the effect of steam on which does not cause changes in their functional properties.
It is used in treatment and prevention and other medical institutions.
Note:
To prevent the formation of scale, increase the efficiency of operation of the
sterilizer it is necessary to use distilled water.
Main technical characteristics
The height of the sterilizer, mm — 1500
The width of the sterilizer, mm — 710
The length of the sterilizer, mm — 1130
The diameter of the wiper cameras, mm — 400
Erase volume cameras, dm3 — 100
Voltage, V — 380
Nominal power, kW — 12
Mass, kg — 230
Manual pressure control
Sterilization modes: 1 mode (tC-min. ) 0.2 MPa — 132-20
2 mode (tC-min. ) 0.1 MPa — 120-45
(//tractor-service.com)
Features:
Economical, simple and easy to use;
Safe for the environment;
The lid, steam jacket and main components are made of high-alloy stainless steel, the heat-protective cover on the lid is made of ABS plastic;
Sterilization chamber made of high-alloy stainless steel brand 12Х18Н10Т (AISI 321);
The camera has the shape of a cylinder and is located horizontally;
Built-in steam generator;
Manually pouring water into the steam generator;
Semi-automatic control;
Heating time — no more than 30 minutes. ;
Preliminary removal of air from the chamber — by the method of gravity purging;
Vacuum drying with the help of a condenser, the depth of vacuum - not less than -0.06 MPa;
Filter for bacterial purification of air entering the chamber at the stage of pressure equalization.
Water consumption for 1 cycle of sterilization (with drying) — no more than 100 l;
The average working time per failure is at least 2,500 cycles;
Service life — 10 years.
Security system:
Shutdown of thermal power plants when the water level in the
steam generator is reduced;
Protective device against overpressure in a steam generator.