Multifunctional portable patient (animal) monitor MEC-1000, Mindray
It is a portable veterinary monitor with numerous functions and is used for monitor monitoring of animals.
The MEC-1000 patient monitor monitors ECG signals, respiratory rate, arterial blood oxygen saturation (SpO2), non-invasive arterial pressure (NIAP) and temperature. The monitor combines parameters measurement modules, a display and a printer in one compact, light and mobile device. Replaceable built-in battery makes patient transportation convenient. A large high-resolution screen provides a clear display of up to 4 waveforms and all basic parameters.
MAIN FEATURES of the patient (animal) monitor MEC-1000, Mindray:
color TFT display 10.4", displaying up to 4 waveforms
modules: ECG, respiration, NIAD, SpO2, temperature, HR
Analysis of arrhythmias, S-T segment for all leads and determination of pacemaker.
Protection against defibrillation and electrosurgical instruments
72 hours of graphic and digital trends of all parameters
tonal modulation of SpO2 pulses (main tone)
sound and visual alarm signaling with a configurable range of activation network connection.
Built-in printer (option)
Internal battery for 1.5 hours of operation.
Animal belongings
Weight: 3.5 kg.
STANDARD SET OF MEC-1000 patient (animal) monitor, Mindray:
modules: ECG, NIAP, respiration, SpO2 (Mindray), temperature (1 channel), arrhythmia analysis, ST-segment analysis, network interface;
accessories: reusable SPO2 sensor for animals with cable, ECG cable with defibrillation protection for 3 leads, NIAD cuff with tube for animals (25-35 cm), built-in rechargeable battery (Lead-Acid 12V).
ADDITIONAL for patient (animal) monitor MEC-1000, Mindray:
modules: thermal printer; accessories: NIAD cuffs of various sizes, temperature sensors, SpO2 sensor, ECG cable.
Technical characteristics
Safety meets the requirements of the IEC60601 series, has CE marking according to MDD93/42/EEC
Dimensions and weight
Dimensions, mm 318x264x152
Weight, kg 4
Terms of use
AC power supply 100-240 V, 50/60 Hz
Temperature 0-40 °С
Humidity 15-95% (without condensation)
Display
Color TFT type, 10.4 inches (diagonal)
Resolution 800x600
Display of 4 waveforms
Scroll speed 6.25 mm/s, 12.5 mm/s, 50 mm/s
Alarm indication, power indication; intermittent QRS sound signal and alarm signaling.
Connection to the network, the possibility of working together with the Mindray Hypervisor central station
Net Port interface
Rechargeable battery Lead-Acid battery (100 minutes of work from two batteries)
Display of trends 1-72 hours
Alarms of 10 levels of sound signaling, visual signaling
The printer is a built-in thermal printer (option)
Printout of channel 2
Print as needed, on alarm, at the established time, etc.
The width of the paper is 50 mm.
Print speed 25mm/s, 50mm/s
ECG
3-lead cable: R, L, F or RA, LF, LL
Registrable referrals I; II; III
Time diagram of single-channel ECG (second channel in cascade)
Choice of gain x0.25; x0.5; x1; x2; auto
Scanning speed 12.5 mm/s, 25 mm/s, 50 mm/s(//tractor-service.com)
Heart rate measurement
Animals 15-300 bpm
Accuracy +/- 1 bpm or +/-1%
Resolution 1 bpm
The protection withstands 4000 V alternating voltage at a frequency of 50 Hz; protection from defibrillation and electrosurgical instruments
The attenuation coefficient of the in-phase signal
Diagnostic mode > 90 dB
Monitoring mode > 105 dB
Surgery mode > 105 dB
Bandwidth 0.05 - 100 Hz
Determining the presence of a pacemaker.
Range of alarms
Animals 15-300 bpm
Analysis of S-T segment availability
The measured interval is -2 mV - 2 mV
Alarm trigger threshold –2 mV - 2 mV
Analysis of arrhythmia presence
Presence of alarms, sound and visual signaling.
Breathing
Impedance measurement method, between R-F (RA-LL)
Measurement range
abdominal 0-150 times/min
Resolution 1 time/min
Error +/- 2 times/min
Range of alarms
animals 0-150 times/min
Alarm of breathing stop 10-40 s
NIAD
Oscillometric measurement method
The measurement mode is automatic/manual/STAT
Selectable unit of measurement: mm. mercury Art. or kPa
Types of measurement of systolic, diastolic and average pressure.
Measurement range
Systolic 40-270 mm Hg. Art.
Diastolic 10-210 mm Hg. Art.
Average 20-330 mm Hg. Art.
Protection from high pressure availability
Resolution 1 mm Hg. Art.
The average error and standard deviation correspond to ANSI/AMI SP-10
Alarms of systolic, diastolic and average pressure
Temperature (central or surface)
The measurement range is 0-50 °С
Resolution 0.1 °С
Error +/-0.1 °С (without sensor error)
Number of channels 1 channel
Alarm range 0-50 °С
SpO2 pulse oximetry
The measurement range is 0-100%
Resolution 1%
Error
at 70-100% +/-2% (adult animals), +/-3% (newborn animals)
at 0-69% undetermined
Alarm range 0-100 %
Pulse rate
Range 25-254 bpm
Resolution 1 bpm
Error +/-2 beats/min or 2%
Alarm range 0-254 bpm
Supported languages Russian, Chinese, English, French, German, Italian, Spanish, Portuguese, Polish, Turkish, Czech, Finnish, Dutch